AATCC TM16 E Color fastness tester exposure conditions
AATCC TM16 E Color fastness tester exposure conditions
One of the AATCC TM16 E test method exposure conditions are as follows.
1. irradiance control at 1.10W/m2/nm@420nm;
2. Window glass filter: continuous light at a black panel temperature of 63°C;
3. Chamber air temperature 43℃, relative humidity 30%;
4. The sample position is kept constant.
Light fastness tester main parameters.
01. Light source: 6500W water-cooled xenon lamp a;
02. Spectral range: full-spectrum daylight;
03. Irradiance closed-loop control range: (307 ~ 1356) W / m2 (300 ~ 800nm) (Boro / Boro inside and outside the filter, other see samples);
04. Irradiance control accuracy: ±0.1W/m2 of the set value;
05. Irradiance intensity uniformity of sample surface: ≦±4%