AATCC TM16 E Color fastness tester exposure conditions

AATCC TM16 E Color fastness tester exposure conditions

One of the AATCC TM16 E test method exposure conditions are as follows.
 
1. irradiance control at 1.10W/m2/nm@420nm;
 
2. Window glass filter: continuous light at a black panel temperature of 63°C;
 
3. Chamber air temperature 43℃, relative humidity 30%;
 
4. The sample position is kept constant.
 
Light fastness tester main parameters.
 
01. Light source: 6500W water-cooled xenon lamp a;
AATCC TM16 E light Color fastness tester
02. Spectral range: full-spectrum daylight;
 
03. Irradiance closed-loop control range: (307 ~ 1356) W / m2 (300 ~ 800nm) (Boro / Boro inside and outside the filter, other see samples);
 
04. Irradiance control accuracy: ±0.1W/m2 of the set value;
 
05. Irradiance intensity uniformity of sample surface: ≦±4%