Analysis of the 4 most common failures of UV Aging Test Chamber
UV Aging Test Chamber common failure analysis:
1, the instrumentation failure rate is high. Fluorescent UV Aging Test Chamber instruments are installed on the top of the equipment, so there is a problem: in the equipment for heating work (sometimes a test will do a few days, or even weeks, the laboratory temperature can reach 60 ℃) and some laboratories do not install air conditioning, and the heat is going up, so the instrument will suffer the ravages of high temperature for a long time, in this high-temperature environment, the instrument is prone to failure.
2,, the water tank heating tube replacement frequently. Because there is no deionized water as required, so the heating tube long-term immersion in tap water, high-temperature tap water. We all know that the heating tube is made of stainless steel tube filled with magnesium powder. So long-term heating work in tap water, will lead to corrosion of the surface of the heating tube stainless steel, rust stains.