UV Aging Test Chamber common failure analysis:

1, the instrumentation failure rate is high. Fluorescent UV Aging Test Chamber instruments are installed on the top of the equipment, so there is a problem: in the equipment for heating work (sometimes a test will do a few days, or even weeks, the laboratory temperature can reach 60 ℃) and some laboratories do not install air conditioning, and the heat is going up, so the instrument will suffer the ravages of high temperature for a long time, in this high-temperature environment, the instrument is prone to failure.

UV Aging Test Chamber

2,, the water tank heating tube replacement frequently. Because there is no deionized water as required, so the heating tube long-term immersion in tap water, high-temperature tap water. We all know that the heating tube is made of stainless steel tube filled with magnesium powder. So long-term heating work in tap water, will lead to corrosion of the surface of the heating tube stainless steel, rust stains.

 