UV Accelerated Weathering Tester common 4 failure analysis
1, UV sink rust through
The sink is used to fill with water for heating, simulating a condensing environment at night in nature. Sink is made of stainless steel, so in the use of equipment, if the user does not follow the prescribed method of adding deionized water to the test, but to join the tap water, will lead to the use of equipment after about 3 years, the formation of substances corroded by water heating, resulting in rusting sink, until the sink and other rust through.

2, ultraviolet lamp point does not light
UV lamps in use, some times the lamp does not light up. This is because the fluorescent UV aging test chamber is generally used are electronic ballasts, electronic ballasts are wired for the jack type. So limited by the details of the production process, it will lead to the lamp corner line in the connection with the electronic ballast there is a hidden danger of insecurity, the use of the gap will occur in the off-line, poor contact and other problems.